PmX-5050 is a fully stand-alone device for continuous analysis of particulate matter (PM) elemental composition, intended for online elemental analysis of heavy metals in the ambient air.
Particulate matter’s presence in the air and its overall accumulation in the environment are the subjects of growing public concern, especially in the manufacturing facilities, plants, and factories surrounding areas. It is vital to measure and assess the PM concentration in the air to prevent any possible harm to natural resources, people residing in such areas and overall environmental deterioration of the region.

Benefits

  • Continuous analysis of PM 2.5 or PM 10 elemental concentration
  • Full compliance with the US EPA IO-3.3 standard
  • Up to 3 months of stand-alone continuous operation
  • Automatic correction of ambient temperature, humidity and pressure influence
  • Heated sampling stage for condensing prevention
  • Simple calibration procedure and tape replacement
  • Intuitive software interface

Previous solutions for PM monitoring were able to measure only overall PM presence and concentration, while full understanding of the PM impact on the human health and environment can only be reached based on elemental analysis.


SPECS

General

Applicable standard: USA EPA IO-3.3

Flow rate: (0~20) L/min

Filter tape: PTFE

Tape replacement interval: up to 3 months

Power supply: AC 100V~240V, 50/60 Hz

Power consumption: 200 W

Dimensions (w/o sampling inlet): 500mm * 500mm * 450mm

Weight (w/o sampling inlet): 40 kg

External connection: USB, Ethernet


XRF unit

X-Ray tube: 50W, 50kV, 3mA

Primary beam filter changer: 5 positions, automatic

X-Ray detector: SDD

Sampling and analysis times: 3 ~ 1440 min

Measurement range: 0 ~ 200 µ g/m3

Detection limit: 0.1 ng/m3 for most elements

Calibration samples: SS316 for energy calibration, NIST 2783 for concentration calibration, other samples(optional)


Environmental Conditions

Ambient temperature: 0 ° C ~ 40 ° C

Humidity: 0 ~ 80%

Altitude: up to 1000m.


Applications

Environmental analysis.

* Attention. Product description can be obtained by machine translation.